MEETINGS / EVENTS

RSS

November 08 - 09 2014, 12:00 AM - 12:00 AM

ASA Quality Meeting 2014

January 23 - 25 2015, 12:00 AM - 12:00 AM

ASA PRACTICE MANAGEMENT 2015

February 07 - 08 2015, 12:00 AM - 12:00 AM

ASA Certificate in Business Administration 2015

>

FDA MEDWATCH ALERTS

October 20, 2014

Lidocaine HCI Injection, USP 10 MG Per ML, 30 ML Single-Dose, Preservative-Free, by Hospira: Recall - Particulate Matter

Summary:

FDA Medwatch Alert 10-20-14

October 16, 2014

FDA MedWatch - LifeCare Flexible Intravenous Solutions by Hospira, Inc.: Recall - Potential for Leakage

Summary:

FDA MedWatch LifeCare Flexible Intravenous Solutions by Hospira Inc

October 13, 2014

FDA MedWatch - CareFusion EnVe and ReVel Ventilators: Class 1 Recall - Power Connection Failure

Summary:

FDA MedWatch CareFusion EnVe and ReVel Ventilators

>

ASA FEATURED PRODUCT

Add this product to your shopping cart

Self-Education and Evaluation (SEE) Program

SKU: 30701-14CE

... Read more »

Single Copies, Member Price: $360

Accutron, Inc. Ultra PC% Cabinet Mount Flowmeters for Nitrous Oxide-Oxygen Sedation Systems: Class I Recall - Flowmeter May Continue To Release Nitrous Oxide Gas When The Oxygen Is Turned Off

Tuesday, October 16, 2012

AUDIENCE: Anesthesiology, Dentistry, Risk Manager

ISSUE: Accutron voluntarily recalled the product after learning through two customer complaints that the flowmeter was flowing nitrous oxide gas without any oxygen gas flow. When not mixed with oxygen, inhaling nitrous oxide can lead to temporary and permanent brain damage and death.

BACKGROUND: The Ultra PC% flowmeter is used to control the flow of gases used in nitrous oxide-oxygen sedation systems. Nitrous oxide-oxygen sedation systems are used to sedate patients during certain dental procedures. The company has reported that two complaints have been received for this flowmeter defect, however, no injuries have been reported to date.

RECOMMENDATION: Consumers with the affected flowmeters should stop using and return the flowmeters to Accutron for a free replacement. The company notified distributors and customers of this recall by mail and is arranging for return and replacement of all recalled products.

Healthcare professionals and patients are encouraged to report adverse events or side effects related to the use of these products to the FDA's MedWatch Safety Information and Adverse Event Reporting Program:

  • Complete and submit the report Online: www.fda.gov/MedWatch/report.htm
  • Download form or call 1-800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed form, or submit by fax to 1-800-FDA-0178

Read the MedWatch alert, including a link to the FDA Recall Notice at:

http://www.fda.gov/Safety/MedWatch/SafetyInformation/SafetyAlertsforHumanMedicalProducts/ucm324120.htm